Hip osteoarthritis is a chronic degenerative disease with usually manifestations after fifty years of life. The method of choice in the treatment of hip osteoarthritis in advanced stages of the disease is total hip arthroplasty. Fitmore® hip stem belongs to the short-tem which saving bone mass in the region of the greater trochanter and the distal part of the femur. In this paper we present the results of total hip arthroplasty  using  Fitmore®   stem in treatment of hip osteoarthritis.

We report a series of 10 patients with osteoarthritis of hip treated by total cementless hip endoprosthesis with Fitmore® Hip Stem at the Clinic for Orthopedics and Traumatology, Clinical Center Nis. The average age of patients was 54.5 (48 - 65) years. There were 5 women and 5 men. Patients were followed for 3 years postoperatively. Functional outcome was assessed according to Harris Hip Score-u. Radiographic analysis was performed on the radiograph of the hip joint in anterior-posterior projection.

In 7 patients we had exellent and in 3 patients  good functional outcome according Harris Hip Score. All patients radiological findings after 3 years shows good integration of Fitmore® stem. The 4 patients had  cortical hypertrophy of the femur.  In patients with hypertrophy of cortex, ther were no clinical manifestations.

 Implantation of total cementless hip joint endoprosthesis with Fitmore®  stem in the treatment of hip osteoarthritis is a good choice in the treatment of younger patients with good bone quality.

References

Aldinger PR, Jung AW, Pritsch M, Breusch S, Thomsen M, Ewerbeck V, et al. Uncemented grit-blasted straight tapered titanium stems in patients younger than fifty-five years of age. Fifteen to twenty-year results. J Bone Joint Surg Am 2009;91(6):1432-39. [CrossRef] [PubMed]

Bensmann G. Cementless fixation of endoprostheses. Biomed Tech 1990;35(Suppl)3:44-77. [CrossRef] [PubMed]

Bobyn JD, Mortimer ES, GlassmanAH, EnghCA, Miller JE, Brooks CE. Producing and avoiding stress shield-ing. Laboratory and clinical observations of non-cemented total hip arthroplasty. Clin Orthop Relat Res 1992;274:79-96. [CrossRef] [PubMed]

Bougherara H, Bureau MN, Yahia L. Bone remodeling in a new biomimetic polymer-composite hip stem. J Biomed Mater Res A 2010;92(1):164-74. [CrossRef] [PubMed]

Garellick G. Swedish Hip Arthroplasty Register. Annual Report 2011. Department of Ortopaedics, Sahlgrenska University Hospital 2012.

Gasbarra E, Celi M, Perrone FL, Iundusi R, Di Primio L, Guglielmi G, et al. Osseointegration of Fitmore stem in total hip arthroplasty. J ClinDensitom 2014; 17(2): 307-13. [CrossRef] [PubMed]

Golubovic Z, Mitkovic M. Degenerativna obolenja zgloba kuka. In: Visnjic M, editor. Hirurgija. Medicinski Fakultet Nis: Prosveta 2013. p.315-6.

Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stemtype femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res 1979;141:17-27. [CrossRef] [PubMed]

Gustke K. Short stems for total hip arthroplasty: initial experience with the Fitmore stem. J Bone Joint Surg Br 2012;94(11 Suppl A): 47-51. [CrossRef] [PubMed]

Hallan G, Lie SA, Furnes O, Engesaeter LB, Vollset SE, Havelin LI. Medium and long-term performance of 11,516 uncemented primary femoral stems from the Norwegian arthroplasty register. J Bone Joint Surg Br 2007;89(12):1574-80. [CrossRef] [PubMed]

Harris WH. Traumatic arthritis of the hip after dis-location and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am. 1969 Jun; 51(4):737-55. [CrossRef] [PubMed]

Maier MW, Streit MR, Innmann MM, Krüger M, Nadorf J, Kretzer JP, et al. Cortical hypertrophy with a short, curved uncemented hip stem does not have any clini-cal impact during early follow-up. BMC Musculoskelet Disord 2015;16:371. [CrossRef] [PubMed]

Pepke W, Nadorf J, Ewerbeck V, Streit MR, Kinkel S, Gotterbam T, et al. Primary stability of the Fimore stem: biomechanical comparison. Int Orthop 2014; 38:483-8. [CrossRef] [PubMed]

Scannell PT, Prendergast PJ. Cortical and interfacial bone changes around a non-cemented hip implant: simulations using a combined strain/damage remo-delling algorithm. Med EngPhys 2009;31(4):477-88. [CrossRef] [PubMed]

Steinhauser E. Biomechanical principles of implant anchoring. In: Gradinger R, Gollwitzer H, editors. Ossear integration. Springer, MedizinVerlag, Heidel-berg; 2006. p.16-23. [PubMed]

Sumner DR, Galante JO. Determinants of stress shielding: design versus materials versus interface. Clin Orthop Relat Res 1992;274:202-12. [CrossRef] [PubMed]

Van Rietbergen B, Huiskes R, Weinans H, Sumner DR, Turner TM, Galante JO. ESB Research Award 1992. The mechanism of bone remodeling and resorption around press- fitted THA stems. J Biomech 1992; 26(4–5):369-82. [PubMed]

Vidalain JP. Twenty-year results of the cementless Corail stem. Int Orthop 2011;35(2):189-94. [CrossRef] [PubMed]

Zimmer, Inc. Fitmore® hip stem - surgical technique. 2009.

Published
2021/01/29
Section
Original article