MVelizarova Evaluation of automated hematology analyzer DYMIND DH76 compared to SYSMEX XN 1000 system

Analytical performance evaluation of DYMIND DH76

  • Milena Velizarova Medical University Sofia, Dpt of Clinical Laboratory
  • Teodora Yacheva Clinical Laboratory & Clinical Pharmacology, UMBAL Alexandrovska, Sofia, Bulgaria
  • Mariana Genova Medical University Sofia, Dpt of Clinical Laboratory
  • Dobrin Svinarov Medical University Sofia, Dpt of Clinical Laboratory
Keywords: hematology analyzer DYMIND DH76, performance evaluation

Abstract


Background: DYMIND DH76 (DYMIND BIOTECH, China) is a new automated hematology system designed to provide CBC count, including 5-part WBC differential count, and its analytical performance should be assessed before adoption for clinical use.

Methods: The analyzer was evaluated according to the International council for standardization in hematology guideline. The purposes of this study were to assess its analytical performance in comparison to SYSMEX XN 1000 hematology analyzer currently used in our laboratory, as well as a comparison between automated and manual WBC differential.

Results: Within-run precision in all concentration ranges was very good with coefficients of variation (CVs) between 0.02% and 2.5% except for platelets over 500×109/L (CV 9.5%). Within-batch imprecision showed CVs lower the declared deviation ranges. Accuracy (defined as trueness) was excellent for all CBC and white cell differential parameters, compared with the state of the art%. Linearity was confirmed with excellent regression coefficients (0.999-1.000), even in the lowest values, and carryover was ≤ 1%. Comparison between DYMIND DH76 and SYSMEX XN 1000 also was very good with correlation coefficients (R2) for WBC (1.000), RBC (0.999), hemoglobin (0.999) and PLT over 50x109/L (0.994) and R2 was lower but still acceptable (0.910) for PLT<50x109/L. R2 for neutrophils, lymphocytes, eosinophils, basophils and monocytes were 0.974, 0.982, 0.957, 0.625 and 0.836, respectively, in the comparison between manual and DYMIND DH76 automated differential WBC counts. Conclusion: With excellent analytical performance and acceptable comparative analysis, DYMIND DH76 hematology analyzer covered the predefined international standards and requirements, and is fully appropriate for clinical application.

References

1. Briggs C, Culp N, Davis B, d'Onofrio G, Zini G, Machin SJ. ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting. Int J Lab Hematol 2014; 36: 613–27.
2. Verbrugge SE, Huisman A. Verification and standardization of blood cell counters for routine clinical laboratory tests. Clin Lab Med 2015; 35: 183–96.
3. Clinical and Laboratory Standards Institute (CLSI). Validation, verification, and quality assurance of automated hematology analyzers. Approved standard. 2nd ed. Wayne, PA: CLSI Document H26-A2: 2010.
4. Vis JY, Huisman A. Verification and quality control of routine hematology analyzers. Int J Lab Hematol 2016; 38: Suppl1: 100-9.
5. Koepke AA, National Committee for Clinical Laboratory Standards. Reference leukocyte (WBC) differential count (proportional) and evaluation of instrumental methods: Approved Standard. H20-A2. Wayne, PA: CLSI; 2007.
6. Marionneaux SM, Maslak PG, Francisco N, Chan V,Hanenberg J, Lynch J,et al. Potential clinical impact of inaccurate automated platelet counts in the setting of severe thrombocytopenia. Blood 2012; 120: 3428;
7. Briggs C, Kunka S, Machin SJ. The most accurate platelet count on the Sysmex XE 2100. Optical or impedance? Clin Lab Haematol2004; 26: 157–158.
8. Buttarello M, Plebani M. Automated blood cell counts: state of the art. Am J Clin Pathol 2008; 130: 104-116.
9. Nakul-Aquaronne D, Sudaka-Sammarcelli I, Ferrero-Vacher C, Starck B, Bayle J. Evaluation of the Sysmex Xe-2100 hematology analyzer in hospital use. J Clin Lab Anal 2003; 17: 113-23.
10. Maciel TS,Comar SR, Beltrame MP.Performance evaluation of the Sysmex® XE-2100D automated hematology analyzer.J Bras Patol Med Lab 2014; 50: 26-35.
11. Kaplan SS, Johnson K, Wolfe N, Brown W, Keeney M, Gray-Statchuk L, Yee IC., et al. Performance characteristics of the Coulter LH 500 hematology analyzer. Lab Hematol 2004; 10: 76-87.
12. McFarlane A, Aslan B, Raby A, Bourner G, Padmore R. Critical values in hematology.Int J Lab Hematol 2015; 37: 36–43.
13. Segal HC, Briggs C, Kunka S, Casbard A, Harrison P, Machin SJ, et al. Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion. Br J Haematol 2005; 128: 520-5.
14. Tanaka Y, Tanaka Y, Gondo K, Maruki Y, Kondo T, Asai S, et al. Performance evaluation of platelet counting by novel fluorescent dye staining in the XN-series automated hematology analyzers. J Clin Lab Anal 2014; 28: 341-8.
15. Bruegel M, Nagel D, Funk M, Fuhrmann P, Zander J, Teupser D. Comparison of five automated hematology analyzers in a university hospital setting: Abbott Cell-Dyn Sapphire, Beckman Coulter DxH 800, Siemens Advia 2120i, Sysmex XE-5000, and Sysmex XN-2000. Clin Chem Lab Med 2015; 53: 1057–71.
16. Becker PH, FenneteauO, Da Costa. Performance evaluation of the Sysmex XN-1000 hematology analyzer in assessment of the white blood cell count differential in pediatric specimens. Int J Lab Hematol 2016; 38: 54-63.
17. Meintker L, Ringwald J, Rauh M, Krause SW. Comparison of automated differential blood cell counts from Abbott Sapphire, Siemens Advia 120, Beckman Coulter DxH 800, and Sysmex XE-2100 in normal and pathologic samples. Am J Clin Pathol 2013; 139: 641–50.
18. Okada T. Development and problem of automated hematology analyzer. Sysmex J Int1999; 9: 52-57.
19. Seo JY, Lee ST, Kim SH. Performance evaluation of the new hematology analyzer Sysmex XN-series. Int J Lab Hematol 2015; 37: 155-64.
20. Vis JY, Huisman A. Verification and quality control of routine hematology analyzers. Int J Lab Hematol. 2016;38 Suppl 1:100-9.
Published
2021/02/18
Section
Original paper