A successful retrieval of stripped outer coating of J-tip diagnostic guidewire from the left popliteal artery during elective coronary angiography

  • Miodrag Damjanović Clinic for Cardiovascular Diseases, Clinical Center of Niš, Niš, Serbia
  • Sonja Šalinger-Martinović Clinic for Cardiovascular Diseases, Clinical Center of Niš, Niš, Serbia
  • Danijela Djordjević-Radojković Clinic for Cardiovascular Diseases, Clinical Center of Niš, Niš, Serbia
  • Goran Koraćević Clinic for Cardiovascular Diseases, Clinical Center of Niš, Niš, Serbia
  • Vladimir Miloradović Clinic for Cardiology, Clinical Center of Kragujevac, Kragujevac, Serbia
Keywords: coronary angiography, intraoperative complications, treatment outcome,

Abstract


Introduction. Entrapment and fracture of diagnostic or therapeutic devices within the coronary circulatory system are a rare, but increasing problem. Case report. A 70-year-old man was admitted in our clinic for coronary angiography before the planned aortic valve replacement. An arterial sheath was inserted in the right common femoral artery. After introducing a J-tip diagnostic coronary guidewire into the aorta and advancing a left Judkins diagnostic catheter over it, suddenly occured peeling off of the wire´s hydrophilic coating at the aortic arch level. Very soon, this outer coating of guidewire carried by the blood stream was entered into the left femoral artery, then into the left popliteal artery. This stripped part of guidewire was successfully caught and extracted out by using a goose-neck snare catheter. Conclusion. A sudden stripping of outer coating of a J-tip diagnostic hydrophilic coronary guidewire during coronary angiography is possible to manage quickly and successfully by the use of a simple cathether.

References

Kaplan S, Kaplan ST, Kutlu M. An unusual case of guide wire fractured during primary percutaneous coronary intervention, and two year follow-up. Kardiol Pol 2010; 68(11): 1291−3.

Goksin I, Baltalarli A, Semiz E, Gurses E, Sacar M, Ozcan V, et al. Catheter entrapment during balloon angioplasty in patient with in-stent restenosis: an unusual complication and its surgical management. J Card Surg 2007; 22(2): 160−2.

Erglis A, Narbute I, Sondore D, Grave A, Jegere S. Tools and tech-niques: coronary guidewires. Euro Intervention 2010; 6: 1−8.

Ammann P, Brunner-la RH, Angehrn W, Roelli H, Sagmeister M, Rickli H. Procedural Complications Following Diagnostic Co-ronary Angiography Are Related to the Operator's Experience and the Catheter Size. Catheter Cardiovasc Interv 2003; 59(1): 13−8.

Hosokawa K, Nakajima Y, Matsuyama H, Shibasaki M. Intraoper-ative monitoring of movement of an entrapped coronary gui-dewire by transesophageal echocardiography. Anesth Analg 2008; 107(4): 1158−60.

Chang TM, Pellegrini D, Ostrovsky A, Marrangoni AG. Surgical management of entrapped percutaneous transluminal coro-nary angioplasty hardware. Tex Heart Inst J 2002; 29(4): 329−32.

Alexiou K, Kappert U, Knaut M, Matschke K, Tugtekin SM. En-trapped coronary catheter remnants and stents: must they be surgically removed. Tex Heart Inst J 2006; 33(2): 139−42.

Karabulut A, Daglar E, Cakmak M. Entrapment of hydrophilic coated coronary guidewire tips: which form of management is best. Cardiol J 2010; 17(1): 104−8.

Criado FJ. Percutaneous arterial puncture and endoluminal access techniques. In: Criado FJ, editor. Endovascular Inter-vention: basic concepts and techniques.1st ed. Oxford: Wiley-Blackwell; 1999. p. 21−9.

Sheth R, Someshwar V, Warawdekar G. Percutaneous retrieval of misplaced intravascular foreign objects with the Dormia basket: an effective solution. Cardiovasc Intervent Radiol 2007; 30(1): 48−53.

Published
2015/04/24
Section
Case report