Usklađenost POCT metode sa potpuno automatizovanom metodom za određivanje HbA1c

  • Neda Milinković Univerzitet u Beogradu-Farmaceutski fakultet, Katedra za medicinsku biohemiju
  • Marija Sarić Matutinović Univerzitet u Beogradu - Farmaceutski fakultet, Katedra za medicinsku biohemiju
  • Gordana Dmitrašinović Univerzitet u Beogradu - Farmaceutski fakultet, Katedra za medicinsku biohemiju
  • Svetlana Ignjatović Univerzitet u Beogradu - Farmaceutski fakultet, Katedra za medicinsku biohemiju; Univerzitetski klinički centar Srbije, Centar za medicinsku biohemiju
Ključne reči: HbA1c, POCT, dijabetes, Demingova regresija

Sažetak


Prethodna istraživanja sugerišu da je „point-of-care“ metoda (POCT) za određivanje glikoziliranog hemoglobina (HbA1c) dijagnostički test koji može biti adekvatna alternativa merenju HbA1c u laboratoriji. Glavni cilj ovog istraživanja bio je da se ispitaju analitičke karakteristike nove INCLIX POCT metode za određivanje HbA1c kako bi se testirale karakteristike izvođenja pre uvođenja ove metode u rutinsku upotrebu. HbA1c je meren u duplikatu u 44 uzorka EDTA krvi paralelno na INCLIX POCT uređaju (Sugitech, Inc.) i automatskim turbidimetrijskim imunoinhibicijskim testom na Olympus AU400 biohemijskom analizatoru (Beckman Coulter). Nepreciznost u seriji bila je 7,58%, nepreciznost između serija bila je 6,63% i 6,22%, a nepreciznost iz dana u dan bila je 8,80% i 7,51%. Ukupna laboratorijska nepreciznost bila je u slaganju sa podacima koje je naveo proizvođač. Izračunat je statistički značajan Pearsonov koeficijent korelacije (r = 0,871, P < 0,01; linearni R2 = 0,757). Demingovom regresionom analizom dobijena je sledeća jednačina: y = -1,80 + 1,304x. Naši rezultati ukazuju na statistički značajnu korelaciju, linearnu povezanost i značajan stepen kompatibilnosti između dve analizirane metode. Međutim, potvrđen je negativan bias vrednosti HbA1c određenih na POCT analizatoru u odnosu na Olympus AU400, što ističe potrebu za standardizacijom INCLIX metode.

Reference

English E, Lenters-Westra E. HbA1c method performance: The great success story of global standardization. Crit Rev Clin Lab Sci. 2018;55(6):408-419.

Lau CS, Aw TC. HbA1c in the diagnosis and management of diabetes mellitus: An update. Diabetes Updates. 2020;6:1-4.

American Diabetes Association. Standards of medical care in diabetes–2021. Diabetes Care. 2021;44(Supplement 1):S15-S33.

Koga M, Okuda M, Inada S, Ueda S, Nakamura Y, Okumiya T, et al. HbA1c levels measured by enzymatic assay during of‑site health checkups are lower than those measured by on‑site HPLC assay. Diabetol Int. 2020;11:67-71.

Arch BN, Blair J, McKay A, Gregory JW, Newland P, Gamble C. Measurement of HbA1c in multicentre diabetes trials – should blood samples be tested locally or sent to a central laboratory: an agreement analysis. Trials 2016;17:517.

Berbudi A, Rahmadika N, Tjahjadi AI, Ruslami R. Performance of Point-of-Care Testing compared with the standard laboratory diagnostic test in the measurement of HbA1c in Indonesian diabetic and nondiabetic subjects. J Diabetes Res. 2020;2020:ID 2037565. doi.org/10.1155/2020/2037565

Jalali MT, Bavarsad SS, Hesam S, Afsharmanesh MR, Mohammadtaghvaei N. Assessing agreement between the three common clinical measurement methods of HbA1c. J Diabetes Metab Disord. 2020;19:273-279.

EurA1c Trial Group. EurA1c: The European HbA1c trial to investigate the performance of HbA1c assays in 2166 laboratories across 17 countries and 24 manufacturers by use of the IFCC model for quality targets. Clin Chem. 2018;64(8):1183-1192.

Standard ISO/TC 212 Clinical laboratory testing and in vitro diagnostic test systems [Internet]. [cited 2022 Jan 07]. Available from: https://www.iso.org/committee/54916.html>

Jeppsson JO, Kobold U, Barr J, Finke A, Hoelzel W, Hoshino T, et al. Approved IFCC reference method for the measurement of HbA(1c) in human blood. Clin Chem Lab Med. 2002;40:78-89.

Approved laboratories of the IFCC network laboratories for HbA1c [Internet]. [cited 2022 Jan 07]. Available from:www.ifcchba1c.net/network/approved

Weykamp C, John WG, Mosca A, Hoshino T, Little R, Jeppsson JO, et al. The IFCC reference measurement system for HbA1c: a 6-year progress report. Clin Chem. 2008;54:240-248.

Erdal GŞ, Işıksaçan N, Koşer M, Kocamaz N. Hemoglobin A1c measurement using Point of Care Testing. Istanbul Med J. 2020;21(1):37-41.

Sølvik US, Røraas TCNG, Sandberg S. Diagnosing diabetes mellitus: Performance of hemoglobin A1c point-of-care instruments in general practice offices. Clin Chem. 2013;59:1790-801.

Arnold WD, Kupfer K, Little RR, Amar M, Horowitz B, Godbole N, et al. Accuracy and precision of a Point-of-Care HbA1c test. J Diabetes Sci Technol. 2020;14(5):883-889.

Clinical and Laboratory Standards Institute. Measurement procedure comparison and bias estimation using patient samples; approved guideline - Third corrected edition. CLSI document EP09-A3. Clinical Laboratory Standards Institute, Wayne, PA, USA, 2018.

Clinical Laboratory Standard Institute. User verification of performance for precision and trueness; approved guideline - Third corrected edition. CLSI document EP15-A2. Clinical and Laboratory Standards Institute, Wayne, PA, USA, 2017.

Clinical and Laboratory Standards Institute. Evaluation of precision performance of quantitative measurement methods; approved guideline - Third corrected edition. CLSI document EP05-A2. Clinical Laboratory Standards Institute, Wayne, PA, USA, 2018.

Australasian Association of Clinical Biochemists. Guidelines for the Evaluation of PoCT Instruments. [Internet]. [cited 2022 Jan 07]. Available:

https://www.aacb.asn.au/documents/item/244

INCLIX HbA1c Test, Instruction for use (2nd revision). Sugentech Inc. Rev.No. IS108E-02/2018.08.01.

Beckman Coulter manufacturer's manual for HbA1c calibrator [Internet]. [cited 2022 Jan 07]. Available: https://www.beckmancoulter.com/wsrportal/techdocs?docname=/cis/BLB00389/%25%25/EN_HbA1c_BLB00389.pdf>

Chesher D. Evaluating assay precision. Clin Biochem Rev. 2008;29:S23-S26.

European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Biological Variation Database. [Internet]. [cited 2022 Feb 02]. Available: https://biologicalvariation.eu/search?q=hba1c>

Knaebel J, Irvin BR, Xie CZ. Accuracy and clinical utility of a point-of-care HbA1c, testing device. Postgrad Med. 2015;125:91-98.

Price CP. Point of care testing. BMJ. 2001;322(7297):1285-1288.

Zhelev Z, Peters J, Rogers M, Andrews R, McDonald T, Hyde C. Accuracy and validity of HbA1c Point of Care Testing: A review of the scientific evidence and guidelines. Exeter Test Group v3. Public Health England, University of Exeter Medical School, 2020.

Hirst JA, McLellan JH, Price CP, English E, Feakins BG, Stevens RJ, et al. Performance of point-of-care HbA1c test devices: aplications for use in clinical practice - a systematic review and meta-analysis. Clin Chem Lab Med. 2017;55:167-180.

Weykamp C. HbA1c: A review of analytical and clinical aspects. Ann Lab Med. 2013;33:393-400.

Heinemann L, Freckmann G. Quality of HbA1c measurement in the practice. The German perspective. J Diabetes Sci Technol. 2015;9(3):687-695.

Ørvim Sølvik U, Risøy AJ, Kjome RLS, Sandberg S. Quality control of Norwegian pharmacy HbA1c testing: a modest beginning. J Diabetes Sci Technol. 2018;12(4):753-61.

Hersberger KE,  Botomino  A,  Mancini  M,  Bruppacher    Sequential screening for diabetes-evaluation of a campaign in Swiss community pharmacies. Pharm World Sci. 2006;28(3):171-179.  

Krass I,  Mitchell  B,  Clarke  P,  Brillant M, Dienaar R, Hughes J, et    Pharmacy diabetes care program: analysis of two screening methods for undiagnosed type 2 diabetes in Australian community pharmacy. Diabetes Res Clin Pract. 2007;75(3):339-347. 

Fikri-Benbrahim N, Martínez-Martínez F, Saéz-Benito L, Luque BS, Corpas JPG, Moullin JC, et al. Assessment of a screening protocol for type 2 diabetes in com-munity   The  DiabNow  Study.  Diabetes  Res  Clin  Pract. 2015;108(3):e49-e52.  

Fornos-Pérez JA,  Andrés-Rodríguez  NF,  Andrés-Iglesias  JC, Luna-Cano R, García-Soidán J, Lorenzo-Veiga B, et al.  Detection  of  people  at  risk  of  diabetes  in  community  pharmacies   of   Pontevedra   (Spain)   (DEDIPO).   Endocrinol Nutr. 2016;63(8):387-396.

Barron E, Misra S, English E, John WG, Sampson M, Bachmann MO, et al. Experience of point-of-care HbA1c testing in the English National Health Service Diabetes Prevention Programme: an observational study. BMJ Open Diab Res Care. 2020;8:e001703.

Van Raalten F, Hiemstra YL, Keulen N, van Duivenvoorde Y, Stoecklein K, Verhagen EA, et al. Level of agreement of point-of-care and laboratory HbA1c measurements in the preoperative outpatient clinic in non-diabetic patients who are overweight or obese. J Clin Monit Comput. 2019;33:1139-1144.

Objavljeno
2022/02/19
Rubrika
Originalni naučni rad